Services Accreditation Blog Contact
Electrical schedule7 min read 22 February 2026

Digital Multimeter Calibration: Resolution or Accuracy?

Reading DMM specifications, solving the “% of reading + % of range” formula, and choosing calibration points on 6½-digit instruments.

In digital multimeter (DMM) calibration, two concepts are often confused: resolution and accuracy. Resolution is the smallest change the instrument can display; accuracy is how close its measurement is to the true value.

The Digit Concept

On DMMs, the term “6½ digits” means the instrument can display 6 full + 1 half (0–1) digit. So the maximum display is 1,999,999. But this is only resolution — it does not indicate accuracy.

The Accuracy Specification Formula

DMM accuracy is usually given as:

Accuracy = ±(% of reading + % of range)

Example: measuring 5 V with a DMM specified at 0.002% of reading + 0.0005% of range on the 10 V range:

Choosing Calibration Points

Typical functions and points in DMM calibration:

Reference Source: Multifunction Calibrator

DMM calibration typically uses multifunction calibrators such as the Fluke 5520A, 5730A or Wavetek 9100. The calibrator’s uncertainty must be at least 4 times smaller than that of the DMM being calibrated (the TAR 4:1 rule).

Tip

Before DMM calibration, leave the instrument switched on for at least 30 minutes at ambient conditions to complete its warm-up. Otherwise the readings will drift.

Resolution or Accuracy?

In practice, resolution is independent of accuracy. The last digit of a 6½-digit DMM may not be meaningful because of noise. Calibration reports include the resolution component in the uncertainty budget using the (d/2)/√3 formula.

Request a quote for DMM and electrical calibration

Service details arrow_forward

Related Articles