Services Accreditation Blog Contact
Quality schedule5 min read 14 February 2026

How to Set the Calibration Interval? An ILAC-G24 Guide

Risk-based interval setting, usage intensity, drift trend and previous calibration data — a practical approach to optimising calibration intervals.

ILAC-G24 (aligned with OIML D 10) is the internationally accepted guidance document for determining and reviewing calibration intervals. ISO/IEC 17025:2017 requires the interval to be justified and set by the laboratory.

Factors Affecting the Interval

ILAC-G24 Methods

The guide defines five different methods for setting the interval:

Practical Starting Values

If there is no historical data, the manufacturer’s recommended interval or industry practice can be a starting point:

Important

The interval is not static. Every calibration result must be analysed; the drift trend must be monitored, and the interval shortened or extended with justification.

Intermediate Checks

To maintain confidence between calibrations, intermediate checks must be performed — for example a repeatability test for a balance, a reference voltage reading for a DMM, or an ice-water point check for a thermometer.

Get consulting for calibration planning

Contact us arrow_forward

Related Articles